Standoff trace explosives vapor detection at meter distances

Abstract

Vapor detection is a noncontact sampling method, which is a less invasive means of explosives screening than physical swiping. Explosive vapor detection is a challenge due to the low levels of vapors available for detection. This study demonstrates that the parts-per-quadrillion sensitivity of atmospheric flow tube-mass spectrometry (AFT-MS) combined with a high-volume air sampler enables standoff detection of trace explosives vapor at distances of centimeters to meters. Standoff detection of explosives vapor was possible both upstream and downstream of the vapor source relative to room air currents. RDX vapor from a saturated source was detected at up to 2.5 m. Vapors from RDX residue and nitroglycerin residue were detected at distances up to 0.5 m. The sampling can be optimized by accounting for air movement in the room or environment, which could further extend standoff detection distances. In conclusion, using AFT-MS with a high-volume sampler could also be effective for standoff vapor detection of drugs and additional chemical threats and could be useful for security screening applications such as at mail facilities, border crossings, and security checkpoints.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Nims, Megan K., Denis, Elizabeth H., Hart, Garret L., Escobedo, Nancy M., Murphy, Shannon E., Addleman, R. Shane, Novosselov, Igor, Ewing, Robert G.. 2023-12-16. Standoff trace explosives vapor detection at meter distances. https://doi.org/10.1016/j.talanta.2023.125562

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related reports

High Multiplicity Trigger for long-lived particles in CMS detector

Searches for long-lived particles (LLPs) at the CMS experiment often involve unconventional event topologies that are difficult to efficiently select using standard trigger strategies. To improve sensitivity to such signatures during LHC Run 3 operation, a dedicated High Multiplicity Trigger (HMT) has been developed and deployed in the CMS trigger system. The trigger targets events containing unusually large numbers of hits in the CMS cathode strip chamber (CSC) muon detectors, a characteristic signature of several LLP scenarios involving displaced decays in the muon system. The HMT implementation, trigger logic, rate dependence with pileup, and operational stability are described. Optimized hit multiplicity thresholds are used to maintain acceptable trigger rates under high-luminosity and high-pileup conditions while preserving high efficiency across a broad range of LLP lifetimes and kinematic regimes. The trigger performance is evaluated using both simulated event samples and proton-proton collision data collected during Run 3 of the LHC. The HMT substantially extends the CMS sensitivity to non-standard signatures associated with LLP decays and provides a flexible platform for future searches for physics beyond the Standard Model.

47 OTHER INSTRUMENTATION↗