Comparison of Detector Based and Source Based Absolute Radiance Standards

Abstract

A detailed comparison of two separate radiometric calibration standards was conducted with analysis of error sources for each. One is a detector based standard, with radiance traceable to fundamental units using the electrical substitution method. The other is based on a calibrated field emission lamp (FEL) generating blackbody radiation. This study was motivated by the discontinuance of the FEL lamps by the manufacturer and a desire to calibrate sensors using non-blackbody spectral profiles. Three spectrometers were calibrated simultaneously by both methods to separate spectrometer artifacts from differences in the radiometric standards and error in the irradiance to radiance conversion procedure needed for the FEL. Conducting this study now is important to provide continuity between the extensive prior FEL based calibration database with a replacement method while calibrated FEL lamps are still available. This also begins a longer term repeatability study of the spectrometers used in this work for evaluation as calibrated transfer standards, usable with both monochromatic and broad spectrum radiance sources. Eliminating the tie to the FEL blackbody spectrum with a detector based standard allows for flexibility in the illumination source. In particular, for Earth science sensors intended for use with sunlit scenes, augmenting traditional tungsten halogen lighting with blue and ultraviolet light emitting diodes allows for a better match to the solar spectrum during laboratory testing.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Brendan McAndrew, Boryana Efremova, Andrei Sushkov, Nathan Kelley, Michael Rodriguez, Nicholas Reed, Julia Barsi. Comparison of Detector Based and Source Based Absolute Radiance Standards. https://ntrs.nasa.gov/citations/20240002383

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related reports

Design, Calibration, and Application of A Cryogenic Low-Background Infrared Radiometer for Spectral Irradiance and Radiance Measurements From 4- to 20-μm Wavelength

We have constructed, calibrated, and tested a cryogenic low-background infrared (IR) radiometer for both spectral radiance and irradiance measurements over the 4- to 20-μm wavelength range. The primary purpose of the Missile Defense Transfer Radiometer (MDXR) is to measure absolute irradiance or radiance from cryogenic IR test chamber sources using a photoconductive Si:As blocked-impurity band (BIB) detector and a set of spectral filters. The MDXR also includes an absolute cryogenic radiometer (ACR) and a Fourier-transform spectrometer (FTS). For irradiance measurements, the ACR is used to provide the primary power scale for the BIB detector in conjunction with spectral filters, whereas the FTS/BIB configuration derives its scale from an internal blackbody source. The two measurement scales show agreement for the irradiance of highly collimated (<1 mrad) IR beams from 10 -13 to 10 -8 W / μm / cm 2 within the combined relative uncertainties of 2.6% (coverage factor k = 1.) We have also calibrated the radiometer for radiance measurements using a large cavity fluid bath blackbody that overfills the spatial and angular extent of the radiometer entrance pupil. The radiometric calibration uncertainty analysis of the radiometer and its maintenance and stability are discussed.

radiometry↗

Spurious Signal in SMAP Fourth Stokes Parameter

The radiometer on the NASA Soil Moisture Active/Passive (SMAP) mission is a fully polarimetric instrument that operates at L-band in the spectrum window at 1400-1427 MHz protected for passive use only. A unique feature of the radiometer is the fully digital back-end which permits direct computation of the third and fourth Stokes parameters: the real and imaginary part of the correlation of signal at horizontal and vertical polarizations, respectively. In particular, the SMAP conical scanning geometry provides the opportunity to look at the global distribution of the fourth Stokes parameter, TA4, at constant incidence angle (40°). A striking feature of TA4 is the existence of a strong (±10 K) spurious signal at coastlines. This article provides examples of the spikes and an explanation of the cause. Simulations have shown that the spurious signal is associated with antenna imperfections, such as cross polarization coupling and phase mismatch between polarizations.

radiometry↗