Search NASA⌕ Search

DOE OSTI · 1632273

Development of Transition-Edge Sensor X-ray Microcalorimeter Linear Array for Compton Scattering and Energy Dispersive Diffraction Imaging

Abstract

In this paper we present a strip transition-edge sensor microcalorimeter linear array detector developed for energy dispersive X-ray diffraction imaging and Compton scattering applications. The prototype detector is an array of 20 transition-edge sensors with absorbers in strip geometry arranged in a linear array. We discuss the fabrication steps needed to develop this array including Mo/Cu bilayer, Au electroplating, and proofof-principle fabrication of long strips of SiNx membranes. We demonstrate minimal unwanted effect of strip geometry on X-ray pulse response and showlinear relationship of 1/pulse height and pulse decay times with absorber length. For the absorber lengths studied, our preliminary measurements show energy resolutions of 40-180eV near 17 keV. Furthermore, we show that the heat flow to the cold bath is nearly independent of the absorber area and depends on the SiNx membrane geometry.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Patel, U., Divan, R., Gades, L., Guruswamy, T., Yan, D., Quaranta, O., Miceli, A.. 2020-02-11. Development of Transition-Edge Sensor X-ray Microcalorimeter Linear Array for Compton Scattering and Energy Dispersive Diffraction Imaging. https://doi.org/10.1007/s10909-019-02267-7

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related reports

High Multiplicity Trigger for long-lived particles in CMS detector

Searches for long-lived particles (LLPs) at the CMS experiment often involve unconventional event topologies that are difficult to efficiently select using standard trigger strategies. To improve sensitivity to such signatures during LHC Run 3 operation, a dedicated High Multiplicity Trigger (HMT) has been developed and deployed in the CMS trigger system. The trigger targets events containing unusually large numbers of hits in the CMS cathode strip chamber (CSC) muon detectors, a characteristic signature of several LLP scenarios involving displaced decays in the muon system. The HMT implementation, trigger logic, rate dependence with pileup, and operational stability are described. Optimized hit multiplicity thresholds are used to maintain acceptable trigger rates under high-luminosity and high-pileup conditions while preserving high efficiency across a broad range of LLP lifetimes and kinematic regimes. The trigger performance is evaluated using both simulated event samples and proton-proton collision data collected during Run 3 of the LHC. The HMT substantially extends the CMS sensitivity to non-standard signatures associated with LLP decays and provides a flexible platform for future searches for physics beyond the Standard Model.

47 OTHER INSTRUMENTATION↗