DOE OSTI · 21266550
ITER LIDAR performance analysis
Abstract
The core LIDAR Thomson scattering for ITER is specified for core profile measurements with a spatial resolution of 7 cm (a/30) for the range of 500 eV<T{sub e}<40 keV and n{sub e}>3x10{sup 19} m{sup -3} at an accuracy of <10% for T{sub e}. These specifications are verified using a full profile Monte Carlo simulation code. In the simulations it is assumed that the input transmission is 50% and the collection transmission is 10% for {lambda}=300-1200 nm and F/=6-17. A crucial design decision lies on the choice of laser and detector combination. It is evaluated that the system can meet its spatial and accuracy specifications for higher temperatures of T{sub e}>5 keV with a combination of a neodymium-doped yttrium aluminum garnet (Nd:YAG) laser ({lambda}{sub 0}=1064 nm, {delta}{lambda}<1 nm, 5 J, and {delta}t{sub FWHM}=250 ps, 5-10 Hz) and S20, GaAs, and GaAsP microchannel plate photomultipliers ({delta}t{sub FWHM}<300 ps, effective quantum efficiency, EQE=3%-4%, and D=18 mm). In order to reach the required T{sub e} of 500 eV with Nd:YAG first harmonic, this choice requires a development of fast near infrared detectors.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Beurskens, M N. A., Kempenaars, M, Scannell, R, Walsh, M J, Giudicotti, L. 2008-10-15. ITER LIDAR performance analysis. https://doi.org/10.1063/1.2969097. https://searchnasa.com/record/osti%3A21266550/iter-lidar-performance-analysis