Abstract

Patient-specific measurements are typically used to validate the dosimetry of intensity-modulated radiotherapy (IMRT). To evaluate the dosimetric performance over time of our IMRT process, we have used statistical process control (SPC) concepts to analyze the measurements from 330 head and neck (H and N) treatment plans. The objectives of the present work are to: (i) Review the dosimetric measurements of a large series of consecutive head and neck treatment plans to better understand appropriate dosimetric tolerances; (ii) analyze the results with SPC to develop action levels for measured discrepancies; (iii) develop estimates for the number of measurements that are required to describe IMRT dosimetry in the clinical setting; and (iv) evaluate with SPC a new beam model in our planning system. H and N IMRT cases were planned with the PINNACLE{sup 3} treatment planning system versions 6.2b or 7.6c (Philips Medical Systems, Madison, WI) and treated on Varian (Palo Alto, CA) or Elekta (Crawley, UK) linacs. As part of regular quality assurance, plans were recalculated on a 20-cm-diam cylindrical phantom, and ion chamber measurements were made in high-dose volumes (the PTV with highest dose) and in low-dose volumes (spinal cord organ-at-risk, OR). Differences between the planned and measured doses were recorded as a percentage of the planned dose. Differences were stable over time. Measurements with PINNACLE{sup 3} 6.2b and Varian linacs showed a mean difference of 0.6% for PTVs (n=149, range, -4.3% to 6.6%), while OR measurements showed a larger systematic discrepancy (mean 4.5%, range -4.5% to 16.3%) that was due to well-known limitations of the MLC model in the earlier version of the planning system. Measurements with PINNACLE{sup 3} 7.6c and Varian linacs demonstrated a mean difference of 0.2% for PTVs (n=160, range, -3.0%, to 5.0%) and -1.0% for ORs (range -5.8% to 4.4%). The capability index (ratio of specification range to range of the data) was 1.3 for the PTV data, indicating that almost all measurements were within {+-}5%. We have used SPC tools to evaluate a new beam model in our planning system to produce a systematic difference of -0.6% for PTVs and 0.4% for ORs, although the number of measurements is smaller (n=25). Analysis of this large series of H and N IMRT measurements demonstrated that our IMRT dosimetry was stable over time and within accepted tolerances. These data provide useful information for assessing alterations to beam models in the planning system. IMRT is enhanced by the addition of statistical process control to traditional quality control procedures.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Breen, Stephen L., Moseley, Douglas J., Zhang, Beibei, Sharpe, Michael B.. 2008-10-15. Statistical process control for IMRT dosimetric verification. https://doi.org/10.1118/1.2975144

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related reports

Deep Learning enabled spectral energy conversion for in situ exposure measurements

A detector-specific deep learning (DL) approach is presented for spectra-to-exposure conversion using large-format sodium iodide (NaI(Tl)) detectors deployed for in situ environmental radiation measurements in emergency response scenarios. Accurate determination of exposure from NaI spectra is challenging due to poor energy resolution, partial energy absorption, and the strong sensitivity of traditionally deployed analytical conversion methods to calibrated source geometry and pre-deployment assumptions. Here, to address these limitations, a multi-layer perceptron model was trained on a hybrid in situ /Monte Carlo dataset constructed to span a broad range of photon energies, spatial extents, and realistic deployment variability, representative of general in situ emergency response conditions. The DL model was evaluated against commonly fielded analytical approaches under matched simulation conditions, including a single-factor method, a G-function method, and a modeled pressurized ion chamber (PIC) baseline. This study was intentionally computational in scope to enable controlled, like-for-like comparisons between conversion techniques while minimizing confounding real-world variability. Comparison to the modeled PIC provides contextual benchmarking and is not intended as a field inter-comparison with deployed instruments. Across the evaluated 20 keV to 3 MeV energy range, the DL approach consistently exhibited higher accuracy and reduced variance relative to the analytical methods against a deterministically calculated exposure. This may indicate improved robustness to spectral complexity without reliance on source-, geometric-, or spectral region-specific optimization. While results do not represent real-world validation, the presented work demonstrates that deep learning may effectively learn the nonlinear detector response-to-exposure relationship for asymmetric NaI(Tl) detectors and offers a promising pathway for improving in situ exposure estimation using spectroscopic systems already integrated into initial real-time emergency response operations.

61 RADIATION PROTECTION AND DOSIMETRY↗

ACRRF High-Bay Dose Calculations using MCNP (Part B)

ACRR radiation outputs through vertical cavities have been documented in two reports. In Part B, the maximum dose from the unshielded central cavity, FREC-II, and NRS is calculated to inform the safety basis. The maximum dose is ≈1430 mrem per 300 MJ, or ≈40,930 rem/hr for full-power operation. A verification study completes the V&V of the modeling. Supplementary studies of variance-reduction techniques, model sensitivities, and aircraft dose above the ACRRF are included.

61 RADIATION PROTECTION AND DOSIMETRY↗

Logistics for the 2026 International Nuclear Accident Dosimetry Intercomparison Exercise

The purpose of this exercise is to test participants’ nuclear accident dosimeters (NADs) and personnel to the requirements established by American National Standards Institute (ANSI), Health Physics Society (HPS) standard ANSI/HPS-N13.3-2013 (R2025) Dosimetry for Criticality Accidents. IER-634 is the second nuclear accident dosimetry exercise taking place at the Armed Forces Radiobiology Research Institute (AFRRI) in Bethesda, MD, a Department of Defense (DOD) facility. This effort continues use of the high-dose neutron exposure test bed maintained by the DOD and represents the continued collaboration between the DOD and Department of Energy (DOE). Previous work at AFRRI included a reactor characterization (IER 484) in 2023 and a nuclear accident intercomparison exercise (IER-602) in 2024.

61 RADIATION PROTECTION AND DOSIMETRY↗